Electrosurgical Analyzer
Foot Switch Control
23
Foot Switch Control
There are three sets of two jacks each on the right side of
the Analyzer. The sets are as follows:
1. FOOT SWITCH (green jacks)
2. FIXED LOAD (white jacks)
3. VAR. LOAD, (red and black jacks)
Note
For the following instructions, refer to the ESU
manufacturer’s instructions to determine the
wiring of the ESU monopolar and bipolar foot
switch jacks. Only qualified technicians should
do this, since the technician will be responsible
for making appropriate and safe test cables.
Refer to Figure 5 and Table 5 for connection examples.
Table 5. Foot Switch Connections
Monopolar Cut
A
Cut activate pin of ESU foot switch
B
Common
Coag
A
Coag activate pin of ESU foot switch
B
Common
Bipolar
A
Bipolar activate pin of ESU foot switch
B
Common