Electrosurgical Analyzer
ESU Tests with the Analyzer
21
Measuring Leakage Current with Grounded HF
Equipment
With the ESU grounded, the test load is 200 ohms, and
the ESU must be operating at maximum power. Test
Setup C as shown in Figure 8 complies with IEC 601.2.2,
sec. 19.101a, test 1, fig. 102, and sec. 19.102, adopted
by ANSI/AAMI HF18-2001.
White
Fixed
Load
Measuring
Load
Black
Earthing
Contact
Red
White
Neutral
Active
QA-ES
ESU
eju013.eps
Figure 8. Test Setup C
Test Setup D shown in Figure 9 complies with IEC
601.2.2, sec. 19.101a, test 2, fig. 103, and sec. 19.102,
adopted by ANSI/AAMI HF18-2001.
White
Fixed
Load
Measuring
Load
Black
Earthing
Contact
Red
White
Neutral
Active
QA-ES
ESU
eju014.eps
Figure 9. Test Setup D
REM Alarm Test
The REM (Return Electrode Monitoring) alarm test
ensures that the ESU sounds an alarm if the resistance
between the two neutral electrodes exceeds its specified