Electrosurgical Analyzer
Analyzer Menu and Function Keys
13
SETUP (F4)
Here you can set the power distribution level for start,
stop, and step in ohms.
F1 F2 F3 F4 F5
eju007.eps
Choose ‘Start load’ by using KNOB PARAM (F4). Use
the encoder to set the level. Save the level by pressing
ENTER (F5). Press CANCEL (F4) to undo an action. Go
to ‘End load’ and ‘Step Size’ and repeat the same
procedure.
Pwr. distr. Start load is the first load during the
measurements; it can be set from 10 ohms to 2100 ohms,
in increments of 25 ohms beginning at 25 ohms.
Pwr. distr. End load is the last load used in the
measurements; it can be set from 525 ohms to 5200
ohms, in increments of 25 ohms from 525 to 2500
ohms and 100 ohms from 2500 ohms to 5200 ohms.
Pwr. distr. Step Size is the load set with steps of 25, 50,
100, and 200 ohms.
QUIT MENU (F5)
Press this key to return to the main menu.
PRINT HEADER (F3)
Press this key to write a heading for a new test protocol.
REMOTE CONTR. (F4)
This selection enables you to control the Analyzer
through a PC, using the Ansur QA-ES Plug-in software
(optional accessory).
Printout
Press PRINT HEADER (F3) before printing out a page if
you want it to have a new heading. The Analyzer
automatically prints out the test results via the printer
output after every measurement. See Figure 3.