PN 4151274 February 2012, Rev. 1 © 2012 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice.
SPOT Light Users Manual iv
v List of Figures Figure Title Page 1. Product Controls and Connections ...
SPOT Light Users Manual vi
1 SpO2 Functional Tester Introduction XW Warning To prevent possible electrical shock, fire, or personal injury, read all “safety information” before
SPOT Light Users Manual 2 Table 1. Symbols Symbol Description W Important information. Refer to manual. X Hazardous voltage. ) Conforms to relevant Ca
SpO2 Functional Tester Unpack the Product 3 • Replace the batteries when the low battery indicator shows to prevent incorrect measurements. • Ca
SPOT Light Users Manual 4 Instrument Familiarization Table 2 is a list of Product controls and connections shown in Figure 1. 21354 gvh003.eps Figure
SpO2 Functional Tester Accessories 5 Accessories Available Product accessories are shown in Table 3 and 4. Table 3. Standard Accessories Item Flu
SPOT Light Users Manual 6 How to Turn On the Product Push for one second to turn on the Product. The screen shown in Figure 2 is the power-up screen
SpO2 Functional Tester How to Use the Product 7 Adjust finger on the display for maximum signal. - Signal + - Signal + -
Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workmanship for two years from the date of o
SPOT Light Users Manual 8 Put the sensor with the LEDs on the bottom of the artificial finger. While you put the sensor on the artifact finger, monito
SpO2 Functional Tester How to Set Custom Tests 9 As one example, to change the SpO2 value to 98 %: 1. Push or to move the highlight to SpO2: i
SPOT Light Users Manual 10 Maintenance The Product is an SpO2 functional tester. Try to prevent mechanical abuse that could change the test values. Th
SpO2 Functional Tester Maintenance 11 How to Clean the Product W Caution Do not put fluid onto the Product surface. Fluid seepage into the electri
SPOT Light Users Manual 12 Battery LED gvh002.eps Figure 6. External Battery Charger Connections
SpO2 Functional Tester Maintenance 13 Battery Removal The battery pack is easy to remove and replace. To remove the battery pack: 1. Push down on
SPOT Light Users Manual 14 General Specifications Temperature Operating ... 10 °C to 40 °C (5
SpO2 Functional Tester Detailed Specifications 15 With Fluke Biomedical R-curves 95 to 100 % ... ±(3
SPOT Light Users Manual 16
Notices All Rights Reserved Copyright 2012, Fluke Biomedical. No part of this publication may be reproduced, transmitted, transcribed, stored in
Returns and Repairs Return Procedure All items being returned (including all warranty-claim shipments) must be sent freight-prepaid to our factory l
Certification This instrument was thoroughly tested and inspected. It was found to meet Fluke Biomedical’s manufacturing specifications when it wa
i Table of Contents Title Page Introduction ...
SPOT Light Users Manual ii Battery Removal ....... 13 G
iii List of Tables Table Title Page 1. Symbols .........................