Operation
Remote Operation
2
2-39
4. Press the softkey corresponding to the parameter to be changed. Repeated
pressing of the softkey cycles through the settings. The following settings
are recommended:
Baud Rate: 300, 600, 1200, 2400, 9600
Parity: None, Even, Odd
Stop Bits: 1 or 2
Data Bits: 7 or 8
The Analyzer factory default setting is 2400, N, 8, 1.
5. Press the esc softkey to store the last displayed parameters in memory.
6. Press esc to return to Main Menu 2.
Ansur Software Control
Ansur test automation systems allow a solutions-based approach to complete
testing of the medical device under test (DUT). Ansur helps you create
standard work using the test template/sequence (which is based on your written
test procedure) and integrates all test results into a single test report that can be
printed or archived. Ansur manages test procedures by allowing both manual
and visual automated test sequences.
The software works hand-in-hand with Fluke Biomedical testers and
simulators, creating a seamless integration for:
Visual inspections
Preventive maintenance
Work procedures
Performance tests
Safety tests
Ansur software utilizes plug-in modules to work with a wide array of Fluke
Biomedical instruments. The plug-in module is a software interface to the
Ansur program. Plug-ins provide test elements used by Ansur. This gives the
benefit of using the same user interface for all testers and simulators supported
by an Ansur plug-in. See the Fluke Biomedical Ansur QED 6 Plug-in User
Manual for detailed information.