IDA-4 Plus
Infusion Pump
Analyzer
Operators
Manual
2242028-IDA4 Plus.qxd 4/22/05 11:08 AM Page 1
Vue de la page 0
1 2 3 4 5 6 ... 73 74

Résumé du contenu

Page 1 - Operators

IDA-4 PlusInfusion PumpAnalyzerOperatorsManual2242028-IDA4 Plus.qxd 4/22/05 11:08 AM Page 1

Page 2

IDA-4 Plus Operators Manual viii 2. Pack the instrument carefully, using the original packing materials if available. Failure to pack the instrumen

Page 3 - Infusion Pump Analyzer

Infusion Pump Analyzer Storage and Shipping ix Storage and Shipping Removing Internal Water before Shipping or Storage Before long-term storage

Page 4

IDA-4 Plus Operators Manual x Warranty Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and workm

Page 5 - Contents

1-1 Chapter 1 Introduction This chapter introduces the instrument and describes its features. Thank you Thank you for purchasing this instrument. I

Page 6

IDA-4 Plus Operators Manual 1-2 Features 9 Tests up to four Infusion Devices simultaneously (when four measuring Transducers fitted) 9 Five but

Page 7

Infusion Pump Analyzer Introduction 1 1-3 Specifications Electrical Specification: Supply Voltage 90 - 260 VAC. Supply Frequency 50 - 60 H

Page 8

IDA-4 Plus Operators Manual 1-4 Physical Dimensions: ~19.05 cm x 18.11 cm x 30.18 (L x W x H) (for rear panel handle add 3.81cm) Weight: 5.0 Kg (w

Page 9 - Notices

2-1 Chapter 2 An Overview This chapter gives an overview of the instrument to help you understand the system. Description of the Device and Intende

Page 10 - Warnings

IDA-4 Plus Operators Manual 2-2 Front Panel Description 1. Liquid Crystal Display-240 dot (W) x 128 (H). Graphic and Alphanumeric. With backlight. 2.

Page 11 - Storage and Shipping

Infusion Pump Analyzer An Overview 2 2-3 Front Panel Layout Drawing channel 1 channel 2inletschannel 3 channel 412345678

Page 13 - Introduction

IDA-4 Plus Operators Manual 2-4 Rear Panel Description 9. Mains ON / OFF switch. 10. Three pin IEC mains inlet connection. 11. Twin fuses integral

Page 14 - Package Contents

Infusion Pump Analyzer An Overview 2 2-5 Rear Panel Layout Drawing

Page 15 - Specifications

IDA-4 Plus Operators Manual 2-6

Page 16 - Physical

3-1 Chapter 3 Operating Instructions This chapter discusses and illustrates the method of connecting units under test and the various operational

Page 17 - An Overview

IDA-4 Plus Operators Manual 3-2 Inlet Hose Connections Drawing

Page 18

Infusion Pump Analyzer Operating Instructions 3 3-3 Outlet Hose Connections Drawing USE ONLY 250V FUSESDRAINAGE TUBESTO FLUID COLLECTIONVESSEL

Page 19 - Front Panel Layout Drawing

IDA-4 Plus Operators Manual 3-4 General Operating Notes The IDA-4 Plus is operated by four directional arrow buttons and an ENTER button on the front

Page 20 - Rear Panel Description

Infusion Pump Analyzer Operating Instructions 3 3-5 MICRO-90 is available from: International Product Corp. 201 Connecticut Dr. P.O. Box 70 Bur

Page 21 - Rear Panel Layout Drawing

IDA-4 Plus Operators Manual 3-6 Start-up Screen When the unit starts, an introductory screen containing the name of the instrument, time, date and fir

Page 22 - Operators Manual

Infusion Pump Analyzer Operating Instructions 3 3-7 Utilities This screen is displayed whenever UTIL is selected. Using the arrow keys to move t

Page 23 - Operating Instructions

IDA-4 Plus Infusion Pump Analyzer Operators Manual April 2005  2005 Fluke Corporation. All rights reserved. Printed in USA All product names are tr

Page 24

IDA-4 Plus Operators Manual 3-8 LCD Set-up Cursor defaults to MODE when this screen is selected, pressing ENTER changes the screen mode to black on wh

Page 25 - TO FLUID COLLECTION

Infusion Pump Analyzer Operating Instructions 3 3-9 Report Header The REPORT HEADER of three lines and up to 28 characters on each line is enter

Page 26 - OPERATING INSTRUCTIONS

IDA-4 Plus Operators Manual 3-10 Engineering Tests Using the arrow keys, move the cursor to the required selection and press ENTER. Test Printer Chec

Page 27 - Warning of Contamination

Infusion Pump Analyzer Operating Instructions 3 3-11 Channel Set-up Menu This screen allows the operator to select the type of test to be perfor

Page 28 - IDA-4 Plus

IDA-4 Plus Operators Manual 3-12 Device Information This screen prompts the operator for device and test information for the specific device under tes

Page 29 - WARNING

Infusion Pump Analyzer Operating Instructions 3 3-13 Sub Menu1 This menu is reached when all the mandatory fields have been entered. From this p

Page 30

IDA-4 Plus Operators Manual 3-14 Location This is an optional field of sixteen characters in length. It is used to store the normal location of the d

Page 31 - RS232 Port

Infusion Pump Analyzer Operating Instructions 3 3-15 Flow Test Screen (Prime Mode) This is the screen that allows the operator to control the fl

Page 32

IDA-4 Plus Operators Manual 3-16 Flow Test Screen (Measuring Mode) When the flow test is started the screen is advanced to the measuring mode as shown

Page 33

Infusion Pump Analyzer Operating Instructions 3 3-17 Flow Test Screen (End of Test Mode) This view of the FLOW TEST screen allows the operator t

Page 34

IDA-4 Plus Operators Manual ii

Page 35

IDA-4 Plus Operators Manual 3-18 Flow Graph Screen The FLOW GRAPH screen shown displays the average flow rate as a continuous line and instantaneous f

Page 36 - Sub Menu2

Infusion Pump Analyzer Operating Instructions 3 3-19 Occlusion Test Screen This is the screen that allows the operator to control the occlusion

Page 37

IDA-4 Plus Operators Manual 3-20 Occlusion Test Screen (Start Mode) When the transducer has completed its tare operation the screen changes as shown.

Page 38 - VOLUME 0.00

Infusion Pump Analyzer Operating Instructions 3 3-21 Occlusion Test Screen (End of Test Mode) This view of the OCCLUSION TEST screen allows the

Page 39 - CHANNEL 1

IDA-4 Plus Operators Manual 3-22 Occlusion Graph Screen This graph displays the pressure value over time. It automatically adjusts its scale so that i

Page 40

Infusion Pump Analyzer Operating Instructions 3 3-23 PCA Test Information Screen This screen is displayed whenever the operator elects to perfor

Page 41 - TIME 00 : 00 : 00

IDA-4 Plus Operators Manual 3-24 bolus will be reported as a loading dose, otherwise a loading dose will not be reported. On completion of this field,

Page 42

Infusion Pump Analyzer Operating Instructions 3 3-25 PCA Test Screen (Start Mode) When the measuring circuit is primed successfully the screen c

Page 43

IDA-4 Plus Operators Manual 3-26 PCA Test Screen (End of Test Mode) This view of the FLOW TEST screen allows the operator to save, print or discard th

Page 44 - Occlusion Graph Screen

Infusion Pump Analyzer Operating Instructions 3 3-27 PCA Graph Screen The PCA GRAPH screen displays the available data in a bar graph format. Th

Page 45

Infusion Pump Analyzer Contents iii Contents Contents...

Page 46

IDA-4 Plus Operators Manual 3-28 PCA Activity with Basal Flow This will follow the PCA Activity Only scenario described above, but with a basal flow r

Page 47

Infusion Pump Analyzer Operating Instructions 3 3-29 Remote Call Input Any change in the Remote Call Input jacks (one for each channel) will be

Page 48

IDA-4 Plus Operators Manual 3-30 Dual Rate Test Information Screen This screen prompts the operator for information required to perform a dual rate fl

Page 49

Infusion Pump Analyzer Operating Instructions 3 3-31 Dual Rate Test Screen This is the screen that allows the operator to control the dual rate

Page 50 - IDA4 Plus

IDA-4 Plus Operators Manual 3-32 Dual Rate Test Screen (Measuring Mode) When the test is started, the screen enters measuring mode for the primary flo

Page 51 - Remote Call Input

Infusion Pump Analyzer Operating Instructions 3 3-33 Dual Rate Test Screen (End of Test Mode) This view of the DUAL FLOW TEST screen allows the

Page 52 - Primary Volume

IDA-4 Plus Operators Manual 3-34 Dual Rate Graph Screen The DUAL FLOW RATE screen displays the average flow rate for the primary and for the secondary

Page 53

Infusion Pump Analyzer Operating Instructions 3 3-35 Measuring Flow-rate against Back-pressure Safety and performance standards for infusion dev

Page 54 - FLOW 1

IDA-4 Plus Operators Manual 3-36 Suggested Apparatus for Back Pressure Testing Drawing chan nel 1 channel 2inletschannel 3 channel 43 - WAY TAPFROM OU

Page 55 - FLOW 2

4-1 Chapter 4 Maintenance, Service and Calibration This chapter discusses routine maintenance and calibration procedures for the user. User Care an

Page 56 - Dual Rate Graph Screen

IDA-4 Plus Operators Manual iv 3 Operating Instructions Connecting Infusion Pumps under Test...

Page 57

IDA-4 Plus Operators Manual 4-2 User Performance Check The Calibration is set at the factory. For accurate Re-Calibration or resetting of Calibration

Page 58

A-1 Appendix A Infusion Devices Delivery Methods This Appendix gives an overview of and discusses different types of Infusion Devices and Delivery

Page 59 - Chapter 4

IDA-4 Plus Operators Manual A-2 Patient Controlled Analgesia (PCA) Pumps These are devices in which the patient is able to initiate a dose of the infu

Page 60

Infusion Pump Analyzer Infusion Devices Delivery Methods A A-3 Control against over-infusion is effective, but increased resistance to flow may

Page 61 - Appendix A

IDA-4 Plus Operators Manual A-4 Infusion Pumps General Common to all pumps is the ability to overcome resistance to flow by increased deliver pressure

Page 62

Infusion Pump Analyzer Infusion Devices Delivery Methods A A-5 automatic alarm and shut down if air enters the system; pre-set control of total

Page 63 - Volumetric Controllers

IDA-4 Plus Operators Manual A-6 A feature of most PCA pumps is a memory log, which can be accessed either through the display, or downloaded via a pri

Page 64 - Volumetric Pumps

Infusion Pump Analyzer Infusion Devices Delivery Methods A A-7 being infused; built-in drug libraries; interface for computer control and monito

Page 65 - PCA Pumps

IDA-4 Plus Operators Manual A-8

Page 66 - Anesthesia Pumps

B-1 Appendix B Computer Control Commands This Appendix gives an overview of computer commands necessary to control the IDA-4 Plus. Overview With

Page 67

Infusion Pump Analyzer Contents v Suggested Apparatus for Back Pressure Testing Drawing ...

Page 68

IDA-4 Plus Operators Manual B-2 Controlling the IDA-4 Plus Using Serial Protocol The IDA-4 Plus allows the operator full flexibility with regard to t

Page 69 - Computer Control Commands

Infusion Pump Analyzer Computer Control Commands B B-3 Perform a Volume Test [CnV,a,b,c] This command starts a flow / volume test on the specif

Page 70 - Command Conventions

IDA-4 Plus Operators Manual B-4 Perform a Dual Flow Rate Test [CnD,a,b,c,d,e,f] This command starts a Dual Flow rate test on the specified channel. T

Page 71

Infusion Pump Analyzer Computer Control Commands B B-5 Fetch the Current Pressure on a Channel [PRESn] Returned to host: [PRES, pppp,hh:mm

Page 72

IDA-4 Plus Operators Manual B-6 ASCII Code Function Number of Parameters IDA-4 Plus < - - > BYE Terminates Computer Control 0 - - > POLL

Page 73

IDA-4 Plus Operators Manual vi

Page 74

Infusion Pump Analyzer Notices vii Notices Fluke Biomedical 6920 Seaway Blvd. Everett, WA 98203 USA Customer Support and Sales USA and Canada: