35080B kVp Divider Operators Manual February 2005 Manual No. 37648 Rev. 7 ©2004, 2005 Fluke Corporation, All rights r
35080B Operators Manual (Blank page)
OperationBattery Check2 2-1 Section 2 Operation Prior to taking measurements with the Model 35080B kVp Divider, it is suggested that the battery
35080B Operators Manual 2-2Figures 2-1A & 2B show suggested oscilloscope settings for the 50-150 kV range using the Wide Range Filter Pack (3761
OperationOscilloscope Zero and Offsets2 2-3Figure 2-1A and 1B. Suggested Oscilloscope Settings for Use with the Wide Range Filter Pack (37617) As
35080B Operators Manual 2-4 Figure 2-2. Geometry of Setup for Above-Table X-ray Tube Figure 2-3. Geometry of Setup for Below-Table X-ray Tube 2.
OperationLinearity Corrections2 2-52.7 Linearity Corrections Corrections for linearity in using the Wide Range Filter Pack (37617) are made by takin
35080B Operators Manual (Blank page)
Theory of OperationIntroduction3 3-1Section 3 Theory of Operation 3.1 Introduction When used with the Wide Range Filter Pack (37617) the Model 35
35080B Operators Manual 3-2 Figure 3-2 Output Ratio Plotted as a Function of kV Figure 3-3. Out
Theory of OperationOutput3 3-33.4 Output In the Model 35080B kVp Divider, the intensity ratio is taken with a high-speed logarithmic ratio module.
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 120 Andrews R
35080B Operators Manual (Blank page)
Test Filter Pack Option (32863)Introduction4 4-1Section 4 Test Filter Pack Option (32863) 4.1 Introduction The Test Filter Pack (32863) provides a
35080B Operators Manual 4-24.4 Operating Instructions The operation of the Test Filter Pack (32863) is similar to the operation of the Wide Range F
Cadmium K-edge Mammo Filter Pack Option (37355)Introduction5 5-1Section 5 Cadmium K-edge Mammo Filter Pack Option (37355) 5.1 Introduction The Cadm
35080B Operators Manual 5-25.2 Specifications Intended Use Beryllium window, molybdenum anode mammographic generators Range 27.5 kVp to 29.5 kVp Ac
Cadmium K-edge Mammo Filter Pack Option (37355)Theory of Operation5 5-3 Figure 5-2. The kVp Correction Curves for the Cadmium
35080B Operators Manual 5-4 2. Oscilloscope Setup - Connect the kVp Divider to the scope using a shielded cable and set the vertical sensitivity s
Cadmium K-edge Mammo Filter Pack Option (37355)Operating Instructions5 5-5Figure 5-4. DC Mammographic Generator Input Waveforms
35080B Operators Manual 5-6 Figure 5-6. Single-phase Mammographic Generator Input Waveform Figure 6-7.
Cadmium K-edge Mammo Filter Pack Option (37355)Operating Instructions5 5-7 When proceeding to Linear Measurement, Section 6.5, Operating Instruc
iTable of Contents Section 1: General Information... 1-1 1.1 Intr
35080B Operators Manual (Blank page)
Linear Mammo Filter Pack Option (37351)Introduction6 6-1Section 6 Linear Mammo Filter Pack Option (37351) 6.1 Introduction The Linear Mammo Filter
35080B Operators Manual 6-2The reading from the Linear Mammo Filter Pack (37351) is compared to that taken with the Cadmium K-edge Mammo Filter Pac
Linear Mammo Filter Pack Option (37351)Corrections6 6-3 Figure 7-2. The correction curve that shows the effect of adding 0.2
35080B Operators Manual 6-4 Figure 6-3. Linear Measurement Setup 2. Connect the kVp Divider output to either a storage oscill
Linear Mammo Filter Pack Option (37351)Operating Instructions6 6-55. Measure other kVp values as desired. Record the LINEAR VOLTAGE, multiply by 1
35080B Operators Manual (Blank page)
CT Filter Pack Option (33551)Introduction7 7-1 Section 7 CT Filter Pack Option (33551) 7.1 Introduction The CT Filter Pack (33551) provides a non-in
35080B Operators Manual 7-27.3 Theory of Operation The Model 35080B kVp Divider contains a pair of solid-state detectors which when exposed to x-ray
CT Filter Pack Option (33551)Correction of Readings7 7-3 Figure 7-1. CT Filter Pack (33551) Filtration Effect Correction Cu
iiSection 6: Linear Mammo Filter Pack Option (37351)... 6-1 6.1 Introduction ...
35080B Operators Manual 7-4 Figure 7-2. CT Filter Pack (33551) Linearity Correction CurvekVCorrection Value (kV) Linearity Corr
Mobile Filter Pack Plus Option (37946)Introduction8 8-1Section 8 Mobile Filter Pack Plus Option (37946) 8.1 Introduction The Mobile Filter Pack Plus
35080B Operators Manual 8-28.2 Specifications Range 50 to 135 kV Sensitivity 0.1 V/10 kV Cal 0.600 V Zero 40 kV Cal 100 kV Calibration Filtration
Mobile Filter Pack Plus Option (37946)Linearity Corrections8 8-38.4 Linearity Corrections Corrections for linearity of the Mobile Filter Pack Plus (3
35080B Operators Manual (Blank page)
Low Range Filter Pack Option (38237)Introduction9 9-1Section 9 Low Range Filter Pack Option (38237) 9.1 Introduction The Low Range Filter Pack (3823
35080B Operators Manual 9-29.4 Operating Instructions The operation of the Low Range Filter Pack (38237) is similar to the operation of the Wide Ran
Low Range Filter Pack Option (38237)Filtration Effect Correction9 9-39.5 Filtration Effect Correction The Low Range Filter Pack (38237) is calibrated
35080B Operators Manual 9-4 Figure 9-2. Low Range Filter Pack (38237) Linearity Correction Curve. The kVp Correction Value
MaintenanceCleaning the Instrument10 10-1 Section 10 Maintenance 10.1 Cleaning the Instrument To clean the exterior of the instrument, follow these
iiiAppendix B: Correcting kVp Measurements for Gain Effects ...B-1 B.1 Introduction ...
35080B Operators Manual 10-2PROBLEMS WITH THE MODEL 35080B KVP DIVIDER 1. The power battery may be weak or dead. Check the battery and replace i
MaintenanceCleaning the Instrument10 10-3 Figure 10-1. Battery Replacement for the Model 35080B kVp Divider 10.4 Calibra
35080B Operators Manual 10-4 Each 1% deviation of this value will result in no more than a 0.75% of reading error. This measurement must be made
MaintenanceCalibration10 10-5 If the oscilloscope in use is not accurate to ± 0.1%, do not attempt adjustment of the Model 35080B kVp Divider. OTHE
35080B Operators Manual 10-6 Figure 10-3. Effects of RP103 Adjustment 10.5 Replaceable Parts Information Fluke Biomedical
MaintenanceCleaning the Instrument10 10-710.6 Recalibration & Repair Information/technical assistance If your instrument needs recalibration or r
35080B Operators Manual (Blank page)
AppendixMammographic Measurements With The Model 35080BA A-1Section A Mammographic Measurements With The Model 35080B A.1 Introduction We recommend
35080B Operators Manual A-2K-EDGE MEASUREMENT PROCEDURE Part A 1. Set the x-ray generator for 25 kVp, 100 mA--or a mAs setting that provides the ma
AppendixLinear MeasurementA A-33. Set scope for single sweep, storage mode, and in the ‘‘DC’’ input mode. LINEAR MEASUREMENT PROCEDURE Part A 1. W
(Blank page)
35080B Operators Manual A-4 Figure A-1. The Model 35080B Mammographic Worksheet (Example) Set kVp K-edge Voltage True kVp Linear Volta
AppendixCorrecting kVp Measurements For Gain EffectsB B-1Section B Correcting kVp Measurements For Gain Effects B.1 Introduction Section 10.4, Cal
35080B Operators Manual B-2B.3 Voltmeter Measurements If the output of the Model 35080B kVp Divider is connected to a voltmeter, a mathematical cor
AppendixThe Model 35050A Dosimeter MeasurementsB B-3 ()offset Zerooffset ZerokVp CAL Actualoffset ZerokVp CAL Expectedoffset ZerokVp MeasuredkVp Co
35080B Operators Manual (Blank page)
AppendixProcedure For Making kVp MeasurementsC C-1Section C Procedure For Making kVp Measurements On Rhodium Anode, Rhodium Filtration Selection Of
35080B Operators Manual C-2Use the following procedure to measure kVp on the rhodium anode, rhodium filtration selection of the DMR. 1. Acquire a C
AppendixProcedureC C-3Table C-2. Look-up-table for Linear Mammo Filter Pack (37351) measurements made on Rh anode, Rh filtration selection of DMR mam
35080B Operators Manual C-44. For all subsequent measurements at other generator settings, use the following procedure. First, obtain a reading us
General InformationIntroduction1 1-1Section 1 General Information 1.1 Introduction The Model 35080B kVp Divider is an alternative to high-voltage
Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564
35080B Operators Manual 1-2 1.4 Specifications Range 50-150 kVp with the Wide Range Filter Pack (37617) Accuracy ± 2% of reading, exclusive of line
General InformationSpecifications1 1-3Table 1-2. The Wide Range Filter Pack (37617) Specifications Range: 50 to 150 kVp Sensitivity: 0.1 V/10 kV C